EM-SA1R

Specifications

Model EM-SA1RP/RS
(base unit)
EM-SA1RP/AN
(base unit)
EM-SA1RC
(sub unit)
LCD Display LCD with 6 digits and polarity, mode display, and two-color
backlight
Resolution 0.1/1/10μm (selectable by setting)
Display range -99.9999∼+99.9999mm
Indication error ±1%±1digit
Response speed Reset / peak clear / hold : Max 20 ms
RS-232C communication : Approx. 2 × n units + 3 ms
Fault detection -NG/OK/+NG/6-rank judgement
RS-232C *1 Equipped Not equipped Not equipped
Analog output *1 Not equipped Equipped Equipped
Number of
detector inputs
1ch
Zero set functino Resetting is possible at any position using a switch on the
front face,an external signal or the RS command.
Data hold method Hold of display and data by external signal
Measuring direction +/-(Can be set switched depending on the setting)
Preset function 6 digits and polarity
Peak measurement +P,-P,P-P,(P-P)/2(+P:Maximum, -P:Minimum)
Error functionality Equipped
Measurement condition registered Seven items (Key, RS command)
Accuracy guaranteed
temperature range
24℃±5℃
Operating ambient temperature 0∼+40℃(to be no dew condensation or freezing)
In storage: -10∼+60℃
Power supply DC12V±10% or DC24V±10%
Power consumption 100mA or less
Weight Approx. 110g Approx. 60g
Dimensions 63(W)×24(H)×48(D)mm
Standards complied with EMC Directive (EN61326-1:2006)

*1 The optional dedicated cable is required. Note that analog output supports absolute value
display.

CLOSE